Gibitre MD Rheometer

17 Nov

Gibitre MD Rheometer

Gibitre MD Rheometer measures the cure
characteristics of a rubber compound in conformity
ty with the international standards.
The measure of the vulcanization is carried out
by measuring the modification in the mechanical characteristics of the sample. The instrument permits to application of a cyclic strain to a test
piece and to measure the associated force. The
test is performed at a defined temperature and
the measure of stiffness is recorded continuously
as a function of time.
Key Features
• Biconical, closed die system, sealed testing
Chamber
• Top brand Torque sensor positioned in the
upper test chamber
• Exclusive construction for the micrometric
adjustment of the gap between the dies
• Independent PID temperature controllers with
0.1¬ ̈‚àûC resolution
• Compressed air cooling circuit for rapid
temperature reduction
• Touch-screen display for instrument control
• Full license of Rheocheck_10 software opti-
mized for Bar-code sample identification
• Full license of Datagest_10 software for com-
plete management of Gibitre SQL Database
Accessories
Automatic sample loader.
Pressure sensor for testing of cellular rubber.
Constant Volume Sample Cutter.
Numerical Test Data: Torque Values: MI, ML, M90, MX, MH, PCR S”@ML,
S”@MH, TanD@ML, TanD@MH. Scorch Time: tS1, tS2, tSX. Cure Time: t90,
tX, tML, tMH, tPCR, tRX, CRI; Pressure: PL, PH, tP, MPR, PR
Displayed Curves: Elastic (S’), Viscose (S”), Complex (S*), Tan-Delta,
storage shear Modulus (G’), loss shear Modulus (G”), Curing speed,
Dies Temperatures
Torque sensor: Brand: Interface®; Capacity: 20 N*m; Resolution: 0.01
dN*m; Linearity Error (%FS): +-0.25
Oscillation frequency: 100 cycles/minute (1,7 ± 0,1 Hz)
Oscillation angle: 0.5°, 1°. Easy adjustment with calibrated gauges
Temperature: between room temperature and +250 °C – Resolution
0.1 °C